JK-90B 900*600*400 Enhanced Version
High Quality Salt Spray Testing Chamber. It's used to test the corrosion resistance of the electrodeposited coating of many kinds of metallic materials.Use the salt spray test machine to spray the Sodium Chloride Solution to the surface of electric coating to achieve the purpose of detection.The material is stainless steel.
1. Product parameter:
Voltage: 110V
Power:2000W
Laboratory volume: 270L
Inner size: 900*600*400mm
Outer size: 1390*830*1100mm
Spray method: continuous
2. Product display:






3. Product information:
-----All equipment needed in the test: Spray nozzle, salt water barrel, test piece support, spray liquid collection container, test chamber, salt water, supply tank, pressure tank, supply equipment and exhaust equipment required for compressed air, the device is shown in the figure, and the test is conducted as follow.
-----Blunt material is required, it is required that the material itself cannot be corroded, and it can not affect the corrosion test.
-----The fog nozzle cannot spray the test liquid directly to the sample , the solution at the top of the spraying chamber cannot drop on the test piece.
-----The test liquid dropping from test piece cannot flow back to the salt bucket again for the test.
-----The solution should not contain grease and dust, so you have to prepare an air cleaner.
-----Spray liquid should be collected for at least 16 hours, and the average of the solution collected in one hour is 1-2ml. Use the final average to represent the amount of spray.
-----The concentration of the Sodium Chloride Solution should be maintained at 40~60g/L.
-----The temperature of the pressure barrel should be kept in 47+1 DEG C, temperature of the salt water barrel should be kept in 35+1 DEG C.
4. Product dimensions:

5. Sample placement:
-----If you need to test many pieces, make sure that all major surfaces can be detected by salt fog at the same time.
-----Arrangement of samples, the spray should be free to drop on all the test pieces, preventing the situation of blocking the spray.
-----Each sample cannot be in contact with each other.
-----Salt spray can’t drop from one test piece to another one.
-----Sample identification mark should be marked at the bottom of the test piece.